Four breakdown solutions introduction of microplate washer

Microplate washer will often appear fault what? To meet these failure need how to deal with?
In the use of various types of washing machine in some common attention, washing machine such as boot check lotion bottle, liquor bottle, water jet needle, suction needle and so on various parts are working properly, if there is a problem should be timely solve.
First, note liquid pipeline not clear
1, the performance of fluid jet strength is insufficient, part of the note liquid pipeline even jam.
2, causes, and treatment
(1) lotion crystallization jam note fluid tube, first off nozzle, reoccupy supporting cleaner poke out tamper.
(2) infusion conduit ageing deformation, joint closes lax, cause leakage, leakage. Should be in duct, tightening joint.
Second, suction fluid pipe not clear
1, performance liquid suction cup after the hole have residual fluid, note after fluid hole in the spill.
2, causes, and treatment
(1) suction fluid tube was fibrin clot, wastepaper and solid jams, processing method to unload nozzle, reoccupy supporting cleaner poke out tamper. Work should be completely and blood coagulation precipitation serum, or with plasma, avoid fiber protein precipitation; Clappers from wastepaper adhesion, anyhow eliminate cutting may cause suction fluid pipe blockage factor.
(2) infusion conduit ageing deformation, joint closes lax, cause leakage, leakage. Should be in duct, tightening joint.
Third, overflow liquid
1, performance imbibing clean but note after fluid lotion overflow cup hole.
2, reasons and processing programming note liquid quantity in cup hole capacity. Modify the program, adjust the filler amount, with the cup of hole capacity phase matching.
Fourth, irrigating needle stuck enzyme label plate (fit eliasa use medical instrument)
1, performance wash the board irrigating needle stuck enzyme label plate, enzyme label plate carrying frame can not continue movement.
2, causes, and treatment
(1) specimen quantity for a little while different project enzyme mark strip in a enzyme label plate rack, each project enzyme mark batten placed rugged, and irrigating needle rising height is certain, high hole stuck irrigating needle. At this time should suspend wash plate, the plate will force higher pore pressure flat.
(2) support lotion needle lift spring aging, stretch inadequacy, flush plate rising height is too low. Should be replaced spring.
As long as the correct use and maintenance of washing machine, appear afore-mentioned small fault promptly eliminate, washing machine will make our enzyme from inspection workstation both relaxed and free from contamination.