Main technical specification of CPAP ventilator machine NLF-200C
1. Environment conditions
Working environment
Ambient temperature 5 ℃~ 40 ℃
Relative humidity ≤80 %
Atmospheric pressure 960 hPa~1040 hPa
Storage and transportation environment
Ambient temperature -10 ℃~+40 ℃
Relative humidity ≤90 %
Atmospheric pressure 860 hPa~1060 hPa
2. Classification
I class,built-in power supply equipment
B class application part
Normal equipment
Continuous operation
Sterilizable(Refer to Chapter 9 of clearance and sterilization)
3. Gas supply
Gas source O2(For medical use)
Air(For medical use)
Rated working pressure 350 kPa
Output pressure of regulator 350 kPa±50 kPa (300 kPa~400 kPa)
4. Power supply
Voltage ~220 V±22 V
Frequency 50 HZ±1 HZ
Input power 900VA
——Main unit 8 W
——Air compressor 670 W
Fuse
——AC power for main unit 5T T1AL 250V
——Auxliary power 5T T1AL 250V
——AC power for air compressor 5T T6.3AL 250V
——AC power for machine 5T T10AL 250V
5. Electromagnetism compatibility
Cell phone and other radiation equipment used near the machine will cause unexpected problem
to the ventilator.
The other equipment added to ventilator will lead to disturbance to the machine. Before used to
the patient, check if the machine works normally under the set values.
6. Parameter adjustment
FiO2
——Adjusting range 21 % ~ 100 %
——Accuracy ± 15 %
CPAP adjusting range 0 cmH2O ~ 15 cmH2O.
Continuous flow adjusting range 2 L/min ~ 20 L/min
7. Parameter monitoring range
FiO2
——Adjusting range 15 % ~ 100 %
——Accuracy ± 5 %
CPAP/airway pressure
——Monitoring range 0 cmH2O ~ 60 cmH2O
——Accuracy ±(2 % full scale + 4 % actual value)
Continuous flow
——Monitoring range 2 L/min ~ 20 L/min
——Accuracy ≥3 L/min:± 20 %
<3 L/min:± 0.6 L/min
8. Technical specification of compressor
Rated output pressure: 0.35 MPa±0.05 MPa(350 kPa±50 kPa)
Continuous output flow ≥25 L/min when output pressure is 0.35MPa
Peak flow: ≥120 L/min
Dew point temperature ≤5 ℃
Alarm for high heat There is sound alarm and light alarm when the compressed
1. Environment conditions
Working environment
Ambient temperature 5 ℃~ 40 ℃
Relative humidity ≤80 %
Atmospheric pressure 960 hPa~1040 hPa
Storage and transportation environment
Ambient temperature -10 ℃~+40 ℃
Relative humidity ≤90 %
Atmospheric pressure 860 hPa~1060 hPa
2. Classification
I class,built-in power supply equipment
B class application part
Normal equipment
Continuous operation
Sterilizable(Refer to Chapter 9 of clearance and sterilization)
3. Gas supply
Gas source O2(For medical use)
Air(For medical use)
Rated working pressure 350 kPa
Output pressure of regulator 350 kPa±50 kPa (300 kPa~400 kPa)
4. Power supply
Voltage ~220 V±22 V
Frequency 50 HZ±1 HZ
Input power 900VA
——Main unit 8 W
——Air compressor 670 W
Fuse
——AC power for main unit 5T T1AL 250V
——Auxliary power 5T T1AL 250V
——AC power for air compressor 5T T6.3AL 250V
——AC power for machine 5T T10AL 250V
5. Electromagnetism compatibility
Cell phone and other radiation equipment used near the machine will cause unexpected problem
to the ventilator.
The other equipment added to ventilator will lead to disturbance to the machine. Before used to
the patient, check if the machine works normally under the set values.
6. Parameter adjustment
FiO2
——Adjusting range 21 % ~ 100 %
——Accuracy ± 15 %
CPAP adjusting range 0 cmH2O ~ 15 cmH2O.
Continuous flow adjusting range 2 L/min ~ 20 L/min
7. Parameter monitoring range
FiO2
——Adjusting range 15 % ~ 100 %
——Accuracy ± 5 %
CPAP/airway pressure
——Monitoring range 0 cmH2O ~ 60 cmH2O
——Accuracy ±(2 % full scale + 4 % actual value)
Continuous flow
——Monitoring range 2 L/min ~ 20 L/min
——Accuracy ≥3 L/min:± 20 %
<3 L/min:± 0.6 L/min
8. Technical specification of compressor
Rated output pressure: 0.35 MPa±0.05 MPa(350 kPa±50 kPa)
Continuous output flow ≥25 L/min when output pressure is 0.35MPa
Peak flow: ≥120 L/min
Dew point temperature ≤5 ℃
Alarm for high heat There is sound alarm and light alarm when the compressed
If you are interested, please contact us per.service003@perlong.com.cn
没有评论:
发表评论